Two-photon fluorescence endoscopy with a micro-optic scanning head

Download (751.63 kB)
journal contribution
posted on 2024-07-13, 08:21 authored by Damian Bird, Min Gu
A major obstacle in the race to develop two-photon fluorescence endoscopy is the use of complicated bulk optics to transmit an ultrashort-pulsed laser beam and return the emitted fluorescence signal. We describe an all-fiber two-photon fluorescence microendoscope based on a single-mode optical fiber coupler, a microprism, and a gradient-index rod lens. It is found that the new endoscope exhibits an axial resolution of 3.2 μm and is capable of imaging transverse cross sections of internal cylindrical structures as small as approximately 3.0 mm in diameter. This device demonstrates the potential for developing a real-time diagnostic tool for biomedical research without the need for surgical biopsy and may find applications in photodynamic therapy, microsurgery, and early cancer detection.

History

Available versions

PDF (Published version)

ISSN

0146-9592

Journal title

Optics Letters

Volume

28

Issue

17

Pagination

2 pp

Publisher

Optical Society of America

Copyright statement

Copyright © 2003 Optical Society of America. The published version is reproduced in accordance with the copyright policy of the publisher.